Standards and Instruments used :
Advance Technology Based high accuracy Standard Digital Cropico Thermometer and Standard/Master Sensors.

Parameters :
Low and high Temperature, Chamber validation

Instruments for calibration :
Contact type temperature sensors, Thermocouple, RTD, and Glass Thermometer, Temperature indicators with sensor, Auto Claves, Incubators, Ovens and Chamber Validation.

Scope of Thermal calibration
S. NoParameter/ measured quantity/ Device Under CalibrationRangeCalibration and Measurement Capability(±)Standard Equipment & Method used
1Temperature Indicator with sensor of Freezer, Chamber, Bath, Oven, & Furnace-50 °C to 200 °C0.2 °C to 0.2 °CUsing Standard RTD, S-Type Sensor with Digital Thermometer by Single Position Calibration (At measuring location in DUC)
2Glass Thermometer, Dial Gauge-40 °C to 200 °C0.65 °C to 0.65 °CUsing Standard RTD Sensor with Standard Digital Thermometer and Low Temperature Bath , By Comparison Method
3Temperature Indicator with sensor of Freezer, Chamber, Bath, Oven, & Furnace600 °C to 1000 °C1.6 °C to 1.6 °CUsing Standard RTD, S-Type Sensor with Digital Thermometer by Single Position Calibration (At measuring location in DUC)
4Temperature Indicator with sensor of Freezer, Chamber, Bath, Oven, & Furnace200 °C to 600 °C0.48 °C to 0.48 °CUsing Standard RTD, S-Type Sensor with Digital Thermometer by Single Position Calibration (At measuring location in DUC)
5Calibration of Contact type temp. sensor (RTDs, Thermocouples, Temp. indicators with Sensors)-40 °C to 100 °C0.11 °C to 0.11 °CUsing Standard RTD, S- Type T/C Sensor with Standard Digital Thermometer and Low Temperature Bath & Dry Block Calibrators , By Comparison Method
6Calibration of Contact type temp. sensor (RTDs, Thermocouples, Temp. indicators with Sensors)100 °C to 200 °C0.14 °C to 0.14 °CUsing Standard RTD, S- Type T/C Sensor with Standard Digital Thermometer and Low Temperature Bath & Dry Block Calibrators , By Comparison Method
7Calibration of Contact type temp. sensor (RTDs, Thermocouples, Temp. indicators with Sensors)200 °C to 600 °C0.32 °C to 0.32 °CUsing Standard RTD, S- Type T/C Sensor with Standard Digital Thermometer and Low Temperature Bath & Dry Block Calibrators , By Comparison Method
8Calibration of Contact type temp. sensor (RTDs, Thermocouples, Temp. indicators with Sensors)600 °C to 1000 °C1.55 °C to 1.55 °CUsing Standard RTD, S- Type T/C Sensor with Standard Digital Thermometer and Low Temperature Bath & Dry Block Calibrators , By Comparison Method
9Calibration of Furnace, Freezer, Oven ,Chamber140 °C to 250 °C1.7 °C to 1.7 °CUsing 16-Channel Temperature Scanner with minimum 9 RTD Sensors by Multi Position Calibration
10Calibration of Furnace, Freezer, Oven ,Chamber-40 °C to 140 °C1.7 °C to 1.7 °CUsing 16-Channel Temperature Scanner with minimum 9 RTD Sensors by Multi Position Calibration

* CMC claimed at best measurement point