Mechanical calibration

Mechanical Calibration

Standards and Instruments Used :
High accuracy Dead weight tester, Advanced Technology based Digital Pressure Gauges and Digital Torque Calibrator, Gauge Block Set K-Grade, Check Master

Parameters :
Negative Pressure, Positive Pressure, Torque, Dimensions

Instruments for Calibration :
Pressure Gauge, Vacuum Gauge, Torque     Screw Driver, Torque Wrench, Plain Plug Gauge, Plain Ring Gauge, Snap gauge. Vernier Caliper,  Micrometer, Gauge Block  (Slip Gauge) up to ‘0’ Grade.

Scope of Mechanical Calibration
S. NoParameter/ measured quantity/ Device Under CalibrationRangeCalibration and Measurement Capability(±)Standard Equipment & Method used
1Calipers (Vernier / Digital /Dial) L.C.: 0.01 mm0 mm to 300 mm13 µm to 13 µmUsing Gauge Block / Caliper Checker by Comparison method (IS 3651 )
2External Micrometer L.C.: 0.001 mm0 mm to 100 mm 2.7 µm to 2.7 µmUsing Gauge Block Set by Comparison method (IS 2967)
3Feeler Gauge0.01 mm to 2 mm2.7 µm to 2.7 µmUsing Digital Micrometer by Comparison method (IS 3179 )
4(Pneumatic–Gauge Pressure) Digital & Dial Pressure Gauge-90 bar to 0 bar0.01 % of rdgUsing Digital Vacuum Gauge & Vacuum Comparator by Comparison Method as per DKD- R - 6 -1
5Hydraulic Pressure # ( Analog / Digital Gauges)0 bar to 700 bar0.1 % of rdgUsing Digital Pressure Gauge with Hydraulic Comparator by Comparison Method as per DKD- R - 6 -1
6(Pneumatic–Gauge Pressure) Digital & Dial Pressure Gauge0 bar to 10 bar0.20 % of rdgUsing Digital Pressure Gauge & Comparator by Comparison Method as per DKD- R - 6 -1
7Hydraulic Pressure ( Analog / Digital Gauges)0.1 bar to 700 bar0.076 % of rdgUsing Dead weight Tester by Comparison Method as per DKD- R - 6 -1
8Force Measuring System of UTM Tension0.2 kN to 10 kN0.59 % to 0.59 %Using Load cell with indicator of Class 0.5 and Class 1 accuracy by UTM of accuracy Class I and coarser based on ISO 1828,Part1
9Force Measuring System of UTM Compression0.2 kN to 10 kN0.24 % to 0.24 %Using Load cell with indicator of Class 0.5 and Class 1 accuracy by UTM of accuracy Class I and coarser based on ISO 1828,Part1
10Glassware like pipettes, burettes, measuring cylinder, volumetric flask etc @ 27°C1 ml to 50 ml0.1 ml to 0.1 mlUsing Weighing balance of 200g capacity and 0.1mg readability and distilled water , Gravimetric method based on IS/ISO 4787
11Glassware like pipettes, burettes, measuring cylinder, volumetric flask etc @ 27°C>50 ml to 100 ml0.5 ml to 0.5 mlUsing Weighing balance of 200g capacity and 0.1mg readability and distilled water , Gravimetric method based on IS/ISO 4787
12Micro-pipette @ 27°C100 µl to 200 µl2.1 µl to 2.1 µlUsing Weighing balance of 200g capacity and 0.1mg readability and distilled water , Gravimetric method based on ISO 8655 part6
13Glassware like pipettes, burettes, measuring cylinder, volumetric flask etc @ 27°C>100 ml to 5000 ml1.33 ml to 1.33 mlUsing Weighing balance of 10kg capacity and 0.1g readability and distilled water ,Gravimetric method based on IS/ISO 4787
14Mass-Electronic weighing balances with readability d=10gMaximum capacity up to 20kg1.2 g to 1.2 gUsing M1 class weights by Calibration of electronic weighing balance and comparator of Class III and coarser as per OIML R-76- 1
15Mass-Electronic weighing balances with readability d=1mgMaximum capacity up to 200g1 mg to 1 mgUsing E2 class weights by Calibration of electronic weighing balance and comparator of Class I and coarser as per OIML R-76- 1
16Mass-Electronic weighing balances with readability d=10g. Maximum capacity up to 100kg12.9 g to 12.9 gUsing M1 class weights by Calibration of electronic weighing balance and comparator of Class III and coarser as per OIML R-76- 1
17Mass-Electronic weighing balances with readability d=10gMaximum capacity up to 60kg3 g to 3 gUsing M1 class weights by Calibration of electronic weighing balance and comparator of Class III and coarser as per OIML R-76- 1
18Mass-Weights5 g to 5 g0.11 mg to 0.11 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
19Mass-Weights10 g to 10 g0.11 mg to 0.11 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
20Mass-Weights1 g to 1 g0.11 mg to 0.11 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
21Mass-Weights2 g to 2 g0.11 mg to 0.11 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
22Mass-Weights20 g to 20 g0.12 mg to 0.12 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
23Mass-Weights50 g to 50 g0.12 mg to 0.12 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
24Mass-Weights100 g to 100 g0.12 mg to 0.12 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
25Mass-Weights1 mg to 1 mg0.11 mg to 0.11 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
26Mass-Weights200 g to 200 g0.16 mg to 0.16 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
27Mass-Weights200 mg to 200 mg0.11 mg to 0.11 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
28Mass-Weights500 mg to 500 mg0.11 mg to 0.11 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
29Mass-Weights100 mg to 100 mg0.11 mg to 0.11 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
30Mass-Weights50 mg to 50 mg0.11 mg to 0.11mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
31Mass-Weights5 mg to 5 mg0.11 mg to 0.11 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
32Mass-Weights2 mg to 2 mg0.11 mg to 0.11 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
33Mass-Weights20 mg to 20 mg0.11 mg to 0.11 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111
34Mass-Weights10 mg to 10 mg0.11 mg to 0.11 mgUsing E2 class weights and balance of readability 0.1mg by Calibration of weights of Class M1 accuracy and coarser as per OIML R-111

* CMC claimed at best measurement point