Optical Calibration

Optical Calibration

Standards and Instruments used :
C-type Goniophotometer with Spectroradiometer, Integrating Sphere with Spectrophotocolorimeter, Standard/Reference Lamps, Spectroradiometer, Illuminance Meter(Multiphotometer).

Parameters :
Luminous Flux, Correlated Colour Temperature, Chromaticity Coordinates, Illuminance(Lux), Luminous Intensity, Wavelength, Irradiance.

Instruments for calibration :
Lamps(CFL,TFL,HPSV,HPMV,MHL,GLS), Illuminance (Lux) Meter, Spectroradiometer, Wavelength(UV Light sources), Irradiance(UV Light sources)

Scope of Optical Calibration
S. NoParameter/ measured quantity/ Device Under CalibrationRangeCalibration and Measurement Capability(±)Standard Equipment & Method used
1Color Coordinates/ HPSV,HPMV, TFL,CFL, MHL ,Tungston filament lampsy =0.001 to y =0.9991.0 % to 1.0 %Using THL lamps at 50W & integrating sphere with standard spectrophoto colorimeter and C-Type mirror Goniometer with Spectroradiometer by Obsolete, ab-initio Method & Substitution method
2Illuminance1 lux to 25000 lux2.1 % to 2.1 %Using Illuminance meter with Intensity Standard Lamp WI 41G Lamp by Direct Method
3Correlated Color Temperature/ HPSV,HPMV, TFL,CFL, MHL ,Tungston filament lamps1500 K to 25000 K1.5 % to 1.5 %Using THL lamps at 50W & integrating sphere with standard spectrophoto colorimeter and C-Type mirror Goniometer with Spectroradiometer by Obsolete, ab-initio Method & Substitution method
4Color Coordinates/ HPSV,HPMV, TFL,CFL, MHL ,Tungston filament lampsx =0.001 to x =0.9991.0 % to 1.0 %Using THL lamps at 50W & integrating sphere with standard spectrophoto colorimeter and C-Type mirror Goniometer with Spectroradiometer by Obsolete, ab-initio Method & Substitution method
5Luminous Intensity1 cd to 10000 cd2.1 % to 2.1 %Using Illuminance meter with Intensity Standard Lamp WI 41G Lamp by Direct Method
6Luminous Flux Of LED Lamp (0.5 W to 35W)20 lm to 3000 lm2.2 % to 2.2 %Using C-Type mirror Goniometer with Spectroradiometer and integrating sphere with standard spectroradiometer/ Spectrophotocolorimeter and Standard light source by Obsolete & Substitution method
7Chromaticity Coordinates of LED Lampx =0.001 to x =0.9991.0 % to 1.0 %Using C-Type mirror Goniometer with Spectroradiometer and integrating sphere with standard spectroradiometer/ Spectrophotocolorimeter and Standard light source by Obsolete & Substitution method
8Correlated Color Temperature of LED Lamp1500 K to 25000 K1.2 % to 1.2 %Using C-Type mirror Goniometer with Spectroradiometer and integrating sphere with standard spectroradiometer/ Spectrophotocolorimeter and Standard light source by Obsolete & Substitution method
9Luminous Flux/ MHL (70W to 400W)4800 lm to 60000 lm3.5 % to 3.5 %Using standard lamps & integrating sphere with standard spectrophoto colorimeter and C-Type mirror Goniometer with Spectroradiometer by Obsolete, ab-initio Method & Substitution method
10Luminous Flux/HPSVL (70W to 400W)4800 lm to 60000 lm3.2 % to 3.2 %Using standard lamps & integrating sphere with standard spectrophoto colorimeter and C-Type mirror Goniometer with Spectroradiometer by Obsolete, ab-initio Method & Substitution method
11Luminous Flux/HPMVL (80W to 400W)2000 lm to 30000 lm3.2 % to 3.2 %Using standard lamps & integrating sphere with standard spectrophoto colorimeter and C-Type mirror Goniometer with Spectroradiometer by Obsolete, ab-initio Method & Substitution method
12Luminous Flux/Compact Fluorescent Lamps (CFL) (5 W to 26 W)200 lm to 1600 lm3.0 % to 3.0 %Using standard lamps with integrating sphere with standard spectrophoto colorimeter and C-Type mirror Goniometer with Spectroradiometer by Obsolete, ab-initio Method & Substitution method
13Luminous Flux/ TFL (6W to 40W)150 lm to 4000 lm3.0 % to 3.0 %Using standard lamps & integrating sphere with standard spectrophoto colorimeter and C-Type mirror Goniometer with Spectroradiometer by Obsolete, ab-initio Method & Substitution method
14Tungsten Filament Lamps (15 W to 200 W)115 lm to 3400 lm3.0 % to 3.0 %Using standard lamps & integrating sphere with standard spectrophoto colorimeter and C-Type mirror Goniometer with Spectroradiometer by ab-initio Method & Substitution method
15Chromaticity Coordinates of LED Lampy =0.001 to y =0.9991.0 % to 1.0 %Using C-Type mirror Goniometer with Spectroradiometer and integrating sphere with standard spectroradiometer/ Spectrophotocolorimeter and Standard light source by Obsolete & Substitution method
16Wavelength/ Spectroradiometer / Radiometer240 nm to 480 nm ((UVC, UVB, UVA and VIS)0.43 nm to 0.99 nmUsing standard Spectroradiometer/Radiometer by comparison method
17Irradiance/ Spectroradiometer / Radiometer 0.1 mW/cm2 to 168 mW/cm2 (UVC, UVB, UVA and VIS) (240 nm to 480 nm)±10 % to ±10 %Using standard Spectroradiometer/Radiometer by comparison method
18Irradiance/ UV Light Source/ Light Source0.1 mW/cm2 to 30 mW/cm2 (UVC, UVB, UVA and VIS) (240 nm to 480 nm)±10 % to  ±10  %Using standard Spectroradiometer/ Radiometer by direct/comparison method
19Wavelength/UV Light Source/ Light Source240 nm to 480 nm (UVC, UVB, UVA and VIS)0.43 nm to 0.99 nmUsing standard Spectroradiometer/ Radiometer by direct/comparison method
20Wavelength/ UV disinfection chamber /UV Chamber240 nm to 480 nm (UVC, UVB, UVA and VIS)0.45 nm to 0.99 nmUsing standard Spectroradiometer/ Radiometer by direct/comparison method
21Irradiance/ UV disinfection chamber /UV Chamber 0.01 mW/cm2 to 30 mW/cm2 (UVC, UVB, UVA and VIS) (240 nm to 480 nm)±10 % to ±10 %Using standard Spectroradiometer/Radiometer by comparison method

* CMC claimed at best measurement point